Software Version C
- 186 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
9.14.6 Switch-on error messages
Immediately after the converter has been switched on, it loads all necessary
parameters, checks configuration and returns to the previous operating mode.
Error messages appearing during this time pop up in one field on the display.
The error number and sometimes a short explanation describe the problem.
To delete the error message, press [CLEAR]. If the error message pops up
again after switching the converter off and on again, a hardware failure has
probably occurred. In this case, please contact our customer service.
Tab. 25 Switch-on error messages and their causes
No. Error message Error messages during switch-on caused by:
1021 INCONSISTENCY IN CONFIGURATION The currently determined hardware configuration does
not match the saved configuration. Check subassembly
monitor and confirm!
1106 ISB CONFLICT Two MA-11 input cards with the same address must not
be installed in the system. Starting the system is
interrupted.
1701 IMPROPER CONFIGURATION
A, B EX?
MA-11 input cards are not allowed to be combined
regarding ex-protection.
Occurs with two MA-11 after exchange.
Check subassembly monitor and confirm!
Operation of ex-sensors not permitted! Contact optek.
1701 IMPROPER CONFIGURATION
C, D EX?
MA-11 input cards are not allowed to be combined
regarding ex-protection.
Occurs with two MA-11 after exchange.
Check subassembly monitor and confirm!
Operation of ex-sensors not permitted! Contact optek.
1701 IMPROPER CONFIGURATION
A, B, C, D EX?
MA-11 input cards are not allowed to be combined
regarding ex-protection.
Occurs with two MA-11 after exchange.
Check subassembly monitor and confirm!
Operation of ex-sensors not permitted! Contact optek.
Tab. 26 Error messages during reading of hardware
No. Error message Error messages during reading of hardware data:
1710 - 1716 INCONSISTENCY IN CALIBRATION DATA
START SYSTEM DIAGNOSTICS
Calibration data on MA-11-1 (input [A], [C], address 1)
optek inputs [A], and [C] not fully functional.
Exchange MA-11 for inputs [A], [C]!
1720 - 1726 INCONSISTENCY IN CALIBRATION DATA
START SYSTEM DIAGNOSTICS
Calibration data on MA-11-2 (input [B], [D], address 2)
optek inputs [B], and [D] not fully functional.
Exchange MA-11 for inputs [B], [D]!
1730 INCONSISTENCY IN CALIBRATION DATA, START
SYSTEM DIAGNOSTICS
Calibration data IO-12, addr. 3 mA-inputs 1, and 2 not
fully functional.
Exchange IO-12!
1740 INCONSISTENCY IN CALIBRATION DATA
START SYSTEM DIAGNOSTICS
Calibration data IO-13, addr. 4 mA-outputs 1, and 2
not fully functional.
Exchange IO-13!
1750 - 1754 INCONSISTENCY IN CALIBRATION DATA
START SYSTEM DIAGNOSTICS
Calibration data IO-14, IO-16, addr. 5 mA-outputs 3,
and 4 not fully functional.
Lamp outputs E, and F not fully functional.
Exchange IO-14 and IO-16!