Software Version C
- 187 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
If errors occur during saving, only one error number is displayed in the error
message field. You must repeat your entry. If a saving error occurs several
times, a hardware failure has probably occurred. In this case, please contact our
customer service.
Tab. 27 Saving error messages and their causes
No. Error message Caused by errors during reading of:
3201 ERROR 3201 Display - contrast, brightness, compensation
3202 ERROR 3202 Format date & time
3203 ERROR 3203 Data logger time interval
3204 ERROR 3204 Damping
3205 ERROR 3205 Lamp voltage E
3206 ERROR 3206 Lamp voltage F
3207 ERROR 3207 Number of sensors
3208 ERROR 3208 Zero point remote / manually
3209 ERROR 3209 Active product after product change
3210 ERROR 3210 mA-outputs 0-20 / 4-20
3211 ERROR 3211 Relay 1-3 limit values / remote
3212 ERROR 3212 Remote In
3213 ERROR 3213 Hold
3214 ERROR 3214 Failsafe -10 % / +110 %
3215 ERROR 3215 Password name
3216 ERROR 3216 Password allocation
3217 ERROR 3217 Password duration
3218 ERROR 3218 optek inputs
3219 ERROR 3219 mA-inputs (assignment)
3220 ERROR 3220 Sensor data
3221 ERROR 3221 TF sensor data
3230 ERROR 3230 Measuring range data
3231 ERROR 3231 Limit values data
3232 ERROR 3232 Display format
3233 ERROR 3233 Product name
3234 ERROR 3234 Linearization
3235 ERROR 3235 Active product
3236 ERROR 3236 Offset + slope set
3240 ERROR 3240 Sensor adaptation
3250 ERROR 3250 Zero point data
3251 ERROR 3251 Configuration