Software Version C
- 115 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
9.10.10 Relay outputs
With this function, you can determine the basic settings for relay outputs. You
can determine whether you want to use the relay outputs 1 to 3 for limit value
indication or status feedback.
The converter is equipped with four relay outputs:
Relay output 1 clamps 10 / 11
Relay output 2 clamps 12 / 13
Relay output 3 clamps 14 / 15
Relay output 4 clamps 16 / 17 / 18 (active)
In this menu, you can switch between the relay outputs 1, 2, and 3. The following
settings are possible:
contacts normal OPEN or normal CLOSED
as outputs for ALARMS (limit value relay) or for Remote Out 1, 2 or 3
(status feedback).
Relay output 4 is a system relay and cannot be switched.
1. Select SYSTEM SETTINGS from the main menu and press [ENTER].
2. Select RELAY OUTPUTS and press [ENTER].
Fig. 118 Menu system settings, choosing relay outputs
SYSTEM SETTINGS
SENSOR TF
mA OUTPUTS
RELAY OUTPUTS
REMOTE IN
HOLD
FAILSAFE
PASSWORD PROTECTION
OPTEK INPUTS
20:05:2006 P01 11:26:55