9.11.8.2.7 Subassembly monitor
With this function, the installed subassemblies can be displayed.
1. Select the SUBASSEMBLY MONITOR function in the MONITORS menu
and press [ENTER].
Fig. 174 Menu service, choosing subassembly monitor
2. The SUBASSEMBLY MONITOR menu appears. The system displays the
exact board names including the revision codes for hardware and software.
If the displays do not correspond to the present subassemblies, a hardware
failure has probably occurred. "Address" signifies internal addresses on the
bus board.
Fig. 175 Subassembly monitor menu
Note!
The "Subassembly monitor menu" figure (fig. 175) shows the assemblies of a
completely equipped converter.
Column "DC" contains variable specifications.
3. Press [CLEAR] to return to the previous menu.
MONITORS
LAMP MONITOR
DETECTOR MONITOR
MEASURING RESULT MONITOR
mA MONITOR
RELAY + LED MONITOR
REMOTE IN MONITOR
SUBASSEMBLY MONITOR
20:05:2006 P01 11:26:55
SUBASSEMBLY MONITOR
BOARD HW SW DC
FRONT PANEL : MC-02 C1
A2 5R0✔
ADDRESS 1 : MA-11-2 A2
A1 5T1✔
ADDRESS 2 : MA-11-2 A2
A1 5T1✔
ADDRESS 3 : IO-12 A2
A2 3R5✔
ADDRESS 4 : IO-13 A2
A3 3R3✔
ADDRESS 5 : IO-14 A2
A3 5T2✔