Software Version C
- 182 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
Tab. 22 Causes and remedies in case of communication errors
No. Communication errors can be caused by: Remedy
1101 Ribbon cable, bus board, plug,
detector card MA-11 (inputs [A], [C], address 1) do not
respond.
Check ribbon cable, bus board, and plug.
Check voltage supply.
Two MA-11 with address 1, exchange card.
Exchange MA-11!
1102 Ribbon cable, bus board, plug,
detector card MA-11 (input [B], [D], address 2) do not
respond.
Check ribbon cable, bus board, and plug.
Check voltage supply.
Two MA-11 with address 2, exchange card.
Exchange MA-11!
1103 Ribbon cable, bus board, plug remote input card IO-12
do not respond.
Check.
Check voltage supply.
Exchange IO-12!
1104 Ribbon cable, bus board, plug relay output card IO-13 do
not respond.
Check ribbon cable, bus board, and plug.
Check voltage supply.
Exchange IO-13!
1105 Ribbon cable, bus board, plug lamp output card IO-14 or
IO-16 do not respond.
Check ribbon cable, bus board, and plug.
Check voltage supply.
Exchange IO-14 or IO-16!
1101
-
1105
Several communication errors occur simultaneously in
case of bus error of main computer.
Check voltage supply.
Exchange MC-02.
Exchange converter!
2011 ERROR
2011
ADC communication error on MA-11-1.
Causes zeroing and calibration errors.
Exchange MA-11 for inputs [A], [C]!
2012 ERROR
2012
ADC communication error on MA-11-2.
Causes zeroing and calibration errors.
Exchange MA-11 for inputs [B], [D]!