EasyManua.ls Logo

Optek C4101 - Page 182

Default Icon
208 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Software Version C
- 172 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
Relay 3: open measuring mode
in MEASUREMENT DISPLAY
in DATA LOGGER DISPLAY
in SYSTEM INFORMATION DISPLAY
closed measurement currently impossible because of
enabled PRODUCT CHANGE (3 sec.)
enabled PRODUCT CONFIGURATION
enabled SYSTEM SETTINGS
enabled MAINTENANCE
enabled HOLD
enabled PC TRANSFER
Note!
In case of SIGNAL LOSS, measuring operation is continued, REL01 and
REL02 remain in the state according to table 15, REL03 remains closed!
During lamp failure, the measuring mode remains active, REL01, REL02
remain in the state according to table 15, REL03 remains closed and
signal loss occurs. The system relay REL04 responds.
You can combine Remote Out 2 with PROD_FEEDBACK (product
feedback via mA-output).
If the relays are defined as status feedback, they switch directly according
to the current status (see above). A delay time for the limit values is not
effected. Set limit values remain valid. It is still indicated by the limit value
LEDs on the front panel when values exceed or fall below the set limit
values.
Relay output 4 (system relay):
The (active) relay responds and switches as follows when switching on the
converter:
Relay 4 (active): open SYSTEM ERROR such as:
no voltage supply
lamp failure
communication error
reading error
configuration error
closed SYSTEM OKAY

Table of Contents