Software Version C
- 152 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
Note!
Switching off the lamp triggers error message 1003 "Failure lamp E", if a lamp
is connected to the system. The LAMP MONITOR menu engages in the system
settings via the user contrary to all other monitors.
Note!
After maintenance switch the lamp outputs on again!
For the LAMP CURRENT menu item, observe the following:
• When lamps fails in the sensors AF16, AS16, AS56, AF26, and TF16-N,
lamp current falls under 0.05 ADC!
• When lamps fails in the sensors AF45 and AF46, lamp current falls to
0.15 - 0.25 ADC if the voltage transformer in the sensor works properly.
• Typical lamp currents for the sensors AF16, AS16, AF26, and TF16-N are
approx. 0.75 ADC!
• Typical lamp currents for the sensors AS56 are approx. 0.43 ADC!
• Typical lamp currents for the sensors AF45, and AF46 are approx.
0.85 ADC!