Software Version C
- 178 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
Tab. 20 Causes and remedies in case of signal loss
Signal loss can be caused by: Remedy
Product translucence is too low.
(Absorbance is too high.)
Check signals in the DETECTOR MONITOR menu.
Check lamp voltage.
Reduce optical path length.
Change wavelength.
Check sensor body window.
Cable break, mixed-up cables Check installation.
Polarity error = signal loss
Lamp failure Check signals in LAMP MONITOR.
Exchange lamp.
Communication error Perform SYSTEM DIAGNOSTICS.
Check SUBASSEMBLY MONITOR.
Exchange faulty boards.