Software Version C
- 180 -
optek-converter C4101 Version 07.2009_1.2US, 23.07.2009
www.optek.com
Tab. 21 Causes and remedies in case of lamp failure
No. Lamp failure can be caused by: Cause Remedy
1001 LAMP CURRENT TOO HIGH Current output E exceeds limit value
(default value: 1.4 A),
short-circuit in lamp cable
Check cable length and
setting.
Check if voltage is applied to
the lamp plug.
Exchange lamp.
Check LAMP MONITOR
menu, exchange IO-14 / IO-16.
1002 LAMP CURRENT TOO HIGH Current output F exceeds limit value
(default value: 1.4 A),
short-circuit in lamp cable
1003 FAILURE LAMP E Current output E falls below limit value
(default value: 0.26 A / 0.60 A, one / two lamps),
broken lamp,
broken lamp output,
cable break,
cables mixed up (UV lamps)
Exchange lamp.
Check high voltage power
supply at UV sensors.
Check if voltage is applied to
the lamp plug.
Check wiring, check polarity.
Check LAMP MONITOR
menu, exchange IO-14 / IO-16.
1004 FAILURE LAMP F Current output F falls below limit value
(default value: 0.26 A / 0.60 A, one / two lamps),
broken lamp,
broken lamp output,
cable break,
cables mixed up (UV lamps)
1005 FAILURE LAMP E Current output E falls below limit value
(default value: 0.26 A),
broken lamp,
broken lamp output,
cable break,
cables mixed up (UV lamps)
1006 FAILURE LAMP F Current output F falls below limit value
(default value: 0.26 A),
broken lamp,
broken lamp output,
cable break,
cables mixed up (UV lamps)
1007 LAMP POWER CONSUMPTION
TOO HIGH
Overload at output E, consumption exceeds
limit value (default value: 15 VA).
wrong lamps connected,
too many lamps connected.
See remedy for error no. 1001,
1002.
Use optek spare parts only.
Check installation.
1008 LAMP POWER CONSUMPTION
TOO HIGH
Overload at output F, consumption exceeds limit
value (default value: 15 VA),
wrong lamps connected,
too many lamps connected.
1011 LAMP POWER CONSUMPTION
TOO HIGH
Overload at output E + F, consumption exceeds
limit value (default value: 22 VA).
1009 FAILURE LAMP E Unacceptable current fluctuations at output E,
lamp voltage too high at UV,
UV lamp with ignition failures,
loose contact with wiring,
collapse of voltage supply, broken lamp output.
Check cable length and lamp
voltage.
Exchange lamp and / or high
voltage power supply.
Check and repair or exchange
wiring.
Check stability, use a line filter.
Check LAMP MONITOR
menu, exchange IO-14 / IO-16.
1010 FAILURE LAMP F Unacceptable current fluctuations at output E,
lamp voltage too high at UV,
UV lamp with ignition failures,
loose contact with wiring,
collapse of voltage supply, broken lamp output.