Spectrum Measurements
R&S
®
ESR
420User Manual 1175.7068.02 ─ 12
Remote command:
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:X on page 868
Peak Search ← Ref Fixed
Sets the maximum value of the selected trace as the reference point.
Remote command:
CALCulate<n>:DELTamarker<m>:FUNCtion:FIXed:RPOint:MAXimum[:PEAK]
on page 867
Marker Demod
The marker demodulation function sends the AM data at the current marker frequency
(in a bandwidth corresponding to the RBW) to the audio output. The "Marker Demod"
softkey opens a submenu to set the demodulation output settings.
For more information see "AF Demodulation" on page 425.
Marker demodulation is not available for Spectrum Emission Mask measurements.
Mkr Demod On/Off ← Marker Demod
Switches the demodulation output on or off.
For more information see "AF Demodulation" on page 425.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation[:STATe] on page 875
AM ← Marker Demod
Sets AM as the output demodulation mode. This is the default setting.
For more information see "AF Demodulation" on page 425.
Remote command:
CALC:MARK1:FUNC:DEM:SEL AM, see CALCulate<n>:MARKer<m>:FUNCtion:
DEModulation:SELect on page 875
FM ← Marker Demod
Sets FM as the output demodulation mode. Default setting is AM.
For more information see "AF Demodulation" on page 425.
Remote command:
CALC:MARK1:FUNC:DEM:SEL FM, see CALCulate<n>:MARKer<m>:FUNCtion:
DEModulation:SELect on page 875
Mkr Stop Time ← Marker Demod
Opens an edit dialog box to define how long demodulation should be output for span >
0.
For more information see "AF Demodulation" on page 425.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:DEModulation:HOLDoff on page 874
Continuous Demod (span > 0) ← Marker Demod
Switches the continuous demodulation on or off. If the sweep time is long enough, the
set frequency range can be monitored acoustically.
Analysis