Spectrum Measurements
R&S
®
ESR
422User Manual 1175.7068.02 ─ 12
Remote command:
CALC:MARK1:FUNC:NDBD:STAT ON, see CALCulate<n>:MARKer<m>:FUNCtion:
NDBDown:STATe on page 878
CALC:MARK1:FUNC:NDBD 3dB, see CALCulate<n>:MARKer<m>:FUNCtion:
NDBDown on page 876
CALC:MARK1:FUNC:NDBD:RES? , see CALCulate<n>:MARKer<m>:FUNCtion:
NDBDown:RESult? on page 877
CALC:MARK:FUNC:NDBD:QFAC?, see CALCulate<n>:MARKer<m>:FUNCtion:
NDBDown:QFACtor on page 877
CALC:MARK1:FUNC:NDBD:FREQ? (span > 0), see CALCulate<n>:MARKer<m>:
FUNCtion:NDBDown:FREQuency? on page 876
CALC:MARK1:FUNC:NDBD:TIME? (span = 0), see CALCulate<n>:MARKer<m>:
FUNCtion:NDBDown:TIME? on page 878
Marker Peak List
Opens the "Peak List" submenu to define criteria for the sort order and the contents of
the peak list. For each listed peak the frequency ("Stimulus") and level ("Response")
values are given. In addition, the peaks are indicated in the trace display. A maximum
of 50 entries are listed.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:COUNt? on page 880
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:X on page 882
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:Y? on page 883
Peak List On/Off ← Marker Peak List
Activates/deactivates the marker peak list. If activated, the peak list is displayed and
the peaks are indicated in the trace display.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:STAT on page 882
Sort Mode Freq/Lvl ← Marker Peak List
Defines the criteria for sorting:
"Freq"
sorting in ascending order of frequency values (span > 0) or time val-
ues (span = 0)
"Lvl"
sorting in ascending order of the level
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SORT on page 882
Max Peak Count ← Marker Peak List
Defines the maximum number of peaks to be determined and displayed.
Remote command:
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE on page 881