Remote Control
R&S
®
ESR
881User Manual 1175.7068.02 ─ 12
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<NoMaxima> Range: 1 to 200
Example:
INIT:CONT OFF
Switches to single sweep mode
INIT;*WAI
Starts measurement and synchronizes to end
CALC:MARK:TRAC 1
Sets marker 1 to trace 1
CALC:MARK:FUNC:FPE:SORT X
Sets the sort mode to increasing X values
CALC:MARK:FUNC:FPE 3;*WAI
Searches the 3 highest maxima for trace 1 and synchronizes to
end
CALC:MARK:FUNC:FPE:COUN?
Queries the number of maxima found
CALC:MARK:FUNC:FPE:Y?
Queries the level of maxima found
CALC:MARK:FUNC:FPE:X?
Queries the frequencies (span <> 0) or time (span = 0) of max-
ima found.
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:LIST:SIZE <MaxNoPeaks>
This command defines the maximum number of peaks the marker peak list may con-
tain.
Suffix:
<n>
.
Selects the measurement window.
<m> Selects the marker.
Parameters:
<MaxNoPeaks> Maximum number of peaks to be determined.
*RST: 50
Example:
CALC:MARK:FUNC:FPE:LIST:SIZE 10
The marker peak list will contain a maximum of 10 peaks.
Manual operation: See "Max Peak Count" on page 422
CALCulate<n>:MARKer<m>:FUNCtion:FPEaks:SEAR:AUTO <State>
This command turns the marker peak search on and off.
This command is retained for compatibility with R&S FSP only. Use CALCulate<n>:
MARKer<m>:FUNCtion:FPEaks:STAT instead.
Remote Commands in Spectrum Analyzer Mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals