CALCulate<n>:LIMit<k>:ACTive?
This command queries the names of all active limit lines.
Suffix:
<n>
.
irrelevant
<k> irrelevant
Return values:
<LimitLines> String containing the names of all active limit lines in alphabeti-
cal order.
Example:
CALC:LIM:ACT?
Queries the names of all active limit lines.
Usage: Query only
CALCulate<n>:LIMit<k>:COMMent <Comment>
This command defines a comment for the selected limit line in all windows.
Suffix:
<n>
.
irrelevant
<k> Selects the limit line.
Parameters:
<Comment> String containing the comment. The maximum number of char-
acters the string may contain is 40.
Example:
CALC:LIM5:COMM 'Upper limit for spectrum'
Defines the comment for limit line 5.
Manual operation: See "Edit Comment" on page 210
CALCulate<n>:LIMit<k>:COPY <Line>
This command copies a limit line.
Suffix:
<n>
.
irrelevant
<k> Selects the limit line.
Parameters:
<Line> 1 to 8
number of the new limit line
<name>
String containing the name of the limit line.
Example:
CALC:LIM1:COPY 2
Copies limit line 1 to line 2.
CALC:LIM1:COPY 'FM2'
Copies limit line 1 to a new line named FM2.
Remote Commands in Receiver Mode

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals