Example:
LIST:RANG2:BAND:RES 5000
Sets the RBW for range 2 to 5 kHz.
Manual operation: See "RBW" on page 290
[SENSe:]LIST:RANGe<range>:BANDwidth:VIDeo
This command sets the VBW value for the specified range.
Suffix:
<range>
.
1...20
range
Parameters:
<Value> Refer to the data sheet.
*RST: 10.0 MHz
Example:
LIST:RANG1:BAND:VID 5000000
Sets the VBW for range 1 to 5 MHz.
Manual operation: See "VBW" on page 290
[SENSe:]LIST:RANGe<range>:BREak <State>
This command configures the sweep behavior.
The suffixes are irrelevant.
Suffix:
<range>
.
1...20
range
Parameters:
<State> ON
The R&S ESR stops after one range is swept and continues only
if you confirm (a message box is displayed).
OFF
The R&S ESR sweeps all ranges in one go.
*RST: OFF
Example:
LIST:RANG:BRE ON
Configures a stop after each range.
Manual operation: See "Stop After Sweep" on page 291
[SENSe:]LIST:RANGe<range>:COUNt
This command returns the number of defined ranges.
Suffix:
<range>
.
1...20
range