Spectrum Measurements
R&S
®
ESR
290User Manual 1175.7068.02 ─ 12
Remote command:
[SENSe:]LIST:RANGe<range>:FILTer:TYPE on page 752
RBW ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the RBW value for this range.
Remote command:
[SENSe:]LIST:RANGe<range>:BANDwidth[:RESolution] on page 750
VBW ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the VBW value for this range.
Remote command:
[SENSe:]LIST:RANGe<range>:BANDwidth:VIDeo on page 751
Sweep Time Mode ← Sweep List dialog box ← Sweep List ← Spurious Emis-
sions
Activates or deactivates the auto mode for the sweep time.
Remote command:
[SENSe:]LIST:RANGe<range>:SWEep:TIME:AUTO on page 757
Sweep Time ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the sweep time value for the range.
Remote command:
[SENSe:]LIST:RANGe<range>:SWEep:TIME on page 757
Detector ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the detector for the range. For details refer to Chapter 6.3.1.5, "Detector Over-
view", on page 399.
Remote command:
[SENSe:]LIST:RANGe<range>:DETector on page 752
Ref. Level ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the reference level for the range.
Remote command:
[SENSe:]LIST:RANGe<range>:RLEVel on page 756
RF Att. Mode ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Activates or deactivates the auto mode for RF attenuation.
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation:AUTO on page 754
RF Attenuator ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Sets the attenuation value for that range.
Remote command:
[SENSe:]LIST:RANGe<range>:INPut:ATTenuation on page 754
Preamp ← Sweep List dialog box ← Sweep List ← Spurious Emissions
Switches the preamplifier on or off.