Remote Control
R&S
®
ESR
902User Manual 1175.7068.02 ─ 12
Example:
INIT:CONT OFF
Selects single sweep operation
CORR:COLL THR;*WAI
Starts the measurement of reference data using direct connec-
tion between generator and device input and waits for the sweep
end.
Usage: Setting only
SCPI confirmed
Manual operation: See "Calibrate Reflection Short" on page 477
See "Calibrate Reflection Open" on page 477
[SENSe:]CORRection:METHod <Type>
This command selects the type of measurement to be performed with the tracking gen-
erator.
The command is available if a tracking generator is installed and has been turned on.
Parameters:
<Type> REFLection
Selects reflection measurements.
TRANsmission
Selects transmission measurements.
*RST: TRANsmission
Example:
CORR:METH TRAN
Sets the type of measurement to "transmission".
Manual operation: See "Calibrate Transmission" on page 477
See "Calibrate Reflection Short" on page 477
See "Calibrate Reflection Open" on page 477
[SENSe:]CORRection:RECall
This command restores the measurement configuration used for calibration.
The command is available if a tracking generator is installed and has been turned on.
Example:
CORR:REC
Usage: Event
SCPI confirmed
Manual operation: See "Recall" on page 478
[SENSe:]CORRection[:STATe] <State>
This command turns correction of measurement results (normalization) on and off.
The command is available after you have defined a reference trace for the selected
measurement type with [SENSe:]CORRection:COLLect[:ACQuire] on page 901.
Remote Commands to Control the Tracking Generator