Remote Control
R&S
®
ESR
903User Manual 1175.7068.02 ─ 12
The command is available if a tracking generator is installed and has been turned on.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
CORR ON
Activates normalization.
Usage: SCPI confirmed
Manual operation: See "Normalize" on page 478
SOURce<n>:AM:STATe <State>
This command turns external amplitude modulation of the tracking generator on and
off.
External I/Q modulation is switched off, if active. This command is available with a
tracking generator (option R&S FSV-B9).
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<State> ON | OFF
*RST: OFF
Example:
SOUR2:AM:STAT ON
Switches on the external amplitude modulation of the tracking
generator for screen B.
Usage: SCPI confirmed
Manual operation: See "External AM" on page 479
See "Modulation OFF" on page 479
SOURce<n>:DM:STATe <State>
This command turns external I/Q modulation of the tracking generator on and off.
External AM and external FM are switched off, if active. This command is available
with a tracking generator (option R&S FSV-B9).
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<State> ON | OFF
*RST: OFF
Remote Commands to Control the Tracking Generator