Remote Control
R&S
®
ESR
901User Manual 1175.7068.02 ─ 12
[SENSe:]CORRection[:STATe].........................................................................................902
SOURce<n>:AM:STATe..................................................................................................903
SOURce<n>:DM:STATe..................................................................................................903
SOURce<n>:FM:DEViation.............................................................................................904
SOURce<n>:FM:STATe.................................................................................................. 904
SOURce<n>:FREQuency:OFFSet................................................................................... 905
SOURce<n>:POWer[:LEVel][:IMMediate][:AMPLitude].......................................................905
SOURce<n>:POWer[:LEVel][:IMMediate]:OFFSet............................................................. 905
SOURce<n>:POWer:MODE............................................................................................906
SOURce<n>:POWer:STARt............................................................................................ 906
SOURce<n>:POWer:STOP.............................................................................................907
OUTPut[:STATe] <State>
This command switches the selected tracking generator on or off.
The command is available with a tracking generator (option R&S FSV-B9).
Parameters:
<State> ON | OFF
*RST: OFF
Example:
OUTP1 ON
Switches on the external tracking generator 1.
Usage: SCPI confirmed
Manual operation: See "Source RF Internal (On/ Off)" on page 477
[SENSe:]CORRection:COLLect[:ACQuire] <MeasType>
This command initiates a reference measurement (calibration). The reference mea-
surement is the basis for the measurement normalization.
To obtain a correct reference measurement, a complete sweep with synchronization to
the end of the sweep must have been carried out. This is only possible in the single
sweep mode.
The command is available if a tracking generator is installed and has been turned on.
Parameters:
<MeasType> THRough
"TRANsmission" mode: calibration with direct connection
between tracking generator and device input
"REFLection" mode: calibration with short circuit at the input
OPEN
only allowed in "REFLection" mode: calibration with open input
Remote Commands to Control the Tracking Generator