EasyManuals Logo

R&S ESR Series User Manual

R&S ESR Series
1126 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #909 background imageLoading...
Page #909 background image
Remote Control
R&S
®
ESR
906User Manual 1175.7068.02 ─ 12
Parameters:
<Offset> Range: -200 dB to +200 dB
*RST: 0dB
Example:
SOUR:POW:OFFS -10dB
Sets the level offset of the tracking generator to – 20 dBm.
Usage: SCPI confirmed
Manual operation: See "Power Offset" on page 486
See "Power Offset" on page 489
SOURce<n>:POWer:MODE
<Mode>
This command turns the power sweep on and off.
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<Mode> FIXed
Power sweep is off.
SWEep
Power sweep is on.
If the power sweep is on, the R&S ESR enters the time domain
(span = 0 Hz). During the sweep time, the power at the internal
tracking generator is changed linearly from start power to stop
power. The start and stop power for the power sweep are dis-
played in the diagram header ("INT TG <start power>… <stop
power>").
*RST: FIXed
Example:
SOUR:POW:MODE SWE
Activates power sweep.
Usage: SCPI confirmed
Manual operation: See "Power Sweep (On /Off)" on page 480
SOURce<n>:POWer:STARt <Power>
This command defines the start power of the power sweep.
Suffix:
<n>
.
1...4
Selects the measurement window.
Parameters:
<Power> Range: -30 dBm to +5 dBm
*RST: 0 dBm
Example:
SOUR:POW:STAR -20dBm
Remote Commands to Control the Tracking Generator

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals