Tracking Generator
R&S
®
ESR
480User Manual 1175.7068.02 ─ 12
Remote command:
SOURce<n>:AM:STATe on page 903
SOURce<n>:DM:STATe on page 903
SOURce<n>:FM:STATe on page 904
Power Sweep
Opens a submenu to define power sweep settings.
This softkey is only available if the (internal) Tracking Generator option (B9) is instal-
led.
This submenu contains the following commands:
●
"Power Sweep (On /Off)" on page 480
●
"Power Sweep Start" on page 480
●
"Power Sweep Stop" on page 480
Power Sweep (On /Off) ← Power Sweep
Activates or deactivates the power sweep. If the power sweep is on, the analyzer is set
to zero span mode (span = 0Hz). During the sweep time of the zero span, the power at
the internal tracking generator is changed linearly from start power to stop power. The
start and stop power for the power sweep are displayed in the diagram header ("INT
TG <start power>… <stop power>", see also Chapter 8.5, "Displayed Information and
Errors", on page 500.
The start power can be set between -30 dBm and +0 dBm.
The stop value can also be be set between -30 dBm and +0 dBm and may be smaller
than the start value.
The difference between the start and stop values may not exceed 10 dB.
This softkey is only available if the (internal) Tracking Generator option (B9) is instal-
led.
Remote command:
SOURce<n>:POWer:MODE on page 906
Power Sweep Start ← Power Sweep
Defines the start power of the power sweep.
The start power can be set between -30 dBm and +0 dBm.
The difference between the start and stop values may not exceed 10 dB.
This softkey is only available if the (internal) Tracking Generator option (B9) is instal-
led.
Remote command:
SOURce<n>:POWer:STARt on page 906
Power Sweep Stop ← Power Sweep
Defines the stop power of the power sweep.
The stop power can be set between -30 dBm and +0 dBm.
The difference between the start and stop values may not exceed 10 dB.
This softkey is only available if the (internal) Tracking Generator option (B9) is instal-
led.
Softkeys of the Tracking Generator Menu