Remote Control
R&S
®
ESR
937User Manual 1175.7068.02 ─ 12
DIAGnostic<n>:SERVice:INPut:PULSed:MCFRequency <Frequency>
This command sets the calibration frequency for frequencies greater than 7 GHz. This
command only takes effect if a microwave calibration signal is selected for input
(DIAGnostic<n>:SERVice:INPut[:SELect] on page 937).
Suffix:
<n>
.
irrelevant
Parameters:
<Frequency> <numeric value>
*RST: 7 GHz
Example:
DIAG:SERV:INP:PULS:MCFR 7,1 GHz
Usage: SCPI confirmed
Manual operation: See "Calibration Frequency MW" on page 527
DIAGnostic<n>:SERVice:INPut[:SELect]
<Signal>
This command selects the source of the signal.
You can feed in a signal from the RF input or the calibration signal.
Suffix:
<n>
.
irrelevant
Parameters:
<Signal> RF
RF input
CALibration
Calibration
*RST: RF
Example:
DIAG:SERV:INP CAL
Selects the calibration signal as the input signal
Manual operation: See "RF" on page 527
DIAGnostic<n>:SERVice:STESt:RESult?
This command queries the self test results.
Suffix:
<n>
.
irrelevant
Parameters:
<Results> String of data containing the results.
The rows of the self test result table are separated by commas.
System Configuration

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the R&S ESR Series and is the answer not in the manual?

R&S ESR Series Specifications

General IconGeneral
BrandR&S
ModelESR Series
CategoryReceiver
LanguageEnglish

Related product manuals