CELL-DYN
®
3/35 Sample Loader TS Guide 9140234B-June 1997
VERIFICATION PROCEDURES
VP - 6.4 Mixing Contact Sensor Check/Adjustment
3 - 97
3
Search
Book TOC
Go Back
7. If
Step 5
or
6
fails, loosen screws on Sensor Bracket and adjust Contact Sensor.
Figure 3-50 [NON-CE MARK] Contact Sensor to Mixing Paddle Clearance
0.010 -0.015 in.
(0.015 Preferred)
(0.254-.381mm)
Sensor
Push Up
With Fin
g
er
Mixin
g
Paddle
SL-200E