6241A/6242 DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-30
Sweep
Sweep range SR0 Auto
SR1 Fixed
SR? Response: SR0 or SR1
Reverse mode SV0 OFF
SV1 ON
SV? Response: SV0 or SV1
Number of times
sweep is repeated
SS cnt
cnt: Count (0
to 1000)
(0 indicates infinite loop)
1
SS? Response: SSdddd
Sweep stop SWSP Stop sweep
Trigger *TRG Sweep start trigger
Measurement trigger
Mea
sure-
ment
Function F0 Measurement OFF
F1 DC Voltage measurement (DCV)
F2 DC current measurement (DCI)
F3 Resistance measurement (OHM)
F?
Response: F0 to
F3
Measurement
range
R0 AUTO range
R1 Fixed to the limiter value range.
(However, if measurement function and source function
are same, range becomes same as source range.)
R? Response: R0 or R1
Measurement
Function Link
Mode
FX0 OFF
FX1 ON (VSIM/ISVM)
(However, if measurement is OFF, measurement
remains OFF after this parameter is changed.)
FX? Response: FX0 or FX1
Item Command Description
Default Operation
Power
ON
Default
setting
During DC/
pulse
operation and
suspension
During
sweep
operation and
suspension