6241A/6242 DC Voltage Current Source/Monitor Operation Manual
6.7.3 Remote Command List
6-33
Cal-
cula-
tion
NULL calculation NL0 OFF
NL1 ON
NL? Response: NL0 or NL1
KNL ± data Sets Null constant (An error occurs if NULL OFF is
set)
*4
0
KNL? Response: KNL ± d.dddddE ± dd
Comparator cal-
culation
CO0 OFF
CO1 ON
CO? Response: CO0 or CO1
KHI ± data Sets upper limit value.
0
KLO ± data Sets lower limit value.
*4
0
KHI? Response: KHI ± d.dddddE ± dd
KLO? KLO ± d.dddddE ± dd
Scaling calcula-
tion
SCL0 OFF
SCL1 ON
SCL? Response: SCL0 or SCL1
KA a a: Constant A (0 (zero) is not available)
1
KB b b: Constant B
0
KC c c: Constant C
*4
1
KA? Response: KA ± d.dddddE ± dd
KB? KB ± d.dddddE ± dd
KC? KC ± d.dddddE ± dd
MAX/ MIN cal-
culation
MN0 OFF
MN1 ON
MN? Response: MN0 or MN1
AVE? Reads out the average value.
0
MAX? Reads out the maximum value.
-9.99999E+26
MIN? Reads out the minimum value.
+9.99999E+26
TOT? Reads out the total value.
0
AVN? Reads out the number of measurement times.
Response: AVN d.dddddE+dd
0
*4: The setting range is from 0 to ± 999.999E + 24.
Item Command Description
Default Operation
Power
ON
Default
setting
During DC/
pulse
operation and
suspension
During
sweep
operation and
suspension