6241A/6242 DC Voltage Current Source/Monitor Operation Manual
2.2.6 Pulse Measurement
2-40
Preparation
1. Follow the same procedure described in Section 2.2.5, “DC Measurement.”
Setting pulse source value
2. Press 123…, 2, UNIT , and ENTER in order.
Set the pulse source value to 2 V.
Setting current-limiter value
3. Press LIMIT, 123…, 3, and ENTER in order.
Set the current-limiter within ±3 mA.
4. Press LIMIT.
The HOME screen is displayed.
Setting base value
5. Press MENU.
6. Press or to select A) SOURCE. Press to go to the Select level.
7. After confirming the item is 1) PLS Base, press to go to the Input/Run level
and press 123…, 1, UNIT , and ENTER in order.
1 V has been set.
Setting pulse time
8. Press , , , , and in this order.
Time setting select level D) TIME is displayed.
9. Press to select 2) Src Delay. Press to go to the Input/Run level.
10. Press 123..., 1, and ENTER in order.
Set the Tds to 1 ms.