CALL:AWGNoise:POWer
CALL:AWGNoise:POWer:STATe[:SELected]
Function Sets/queries the additive white gaussian noise (AWGN) state.
The optional [:SELected] keyword in this command specifies that the AWGN power level being
queried applies to the current system type.
AWGN and cell power (see CALL[:CELL]:POWer ) are absolute power levels and make up the
total forward channel RF output power. Depending on the cell power setting, an AWGN level may
be within the range listed below but cause an error message to be displayed. Possible causes are:
Sum of AWGN and cell power causing total RF power limits to be exceeded (see
CALL:TOTal:POWer[:AMPLitude][:SELected]? for range of total RF power.)
Greater than 10 dB difference between cell power and AWGN.
Amplitude offset causing total RF power limits to be exceeded (see
SYSTem:CORRection:GAIN to set or query the amplitude offset value).
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting OFF
Programming Example
OUTPUT 714;"CALL:AWGNOISE:POWER:STATE:SELECTED ON" !Turns on AWGN.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals