CALL[:CELL]:POWer
CALL[:CELL]:POWer:STATe[:SELected]
Function Sets/queries the cell power state.
The optional [:SELected] keyword in this command specifies that the cell power state being set or
queried applies to the current system type. The current system type can be changed using the
CALL[:CELL]:SYSTem[:TYPE] command.
Cell power and AWGN (see CALL:AWGNoise:POWer ) are absolute power levels and make up
the total forward channel RF output power. Depending on the cell power setting, a cell power
level may be within the range listed below but cause an error message to be displayed. Possible
causes are:
Sum of cell power and AWGN causing total RF power limits to be exceeded (see
CALL:STATus:TOTal:POWer[:AMPLitude][:SELected]? for the allowable range of total RF
power levels.)
Greater than 10 dB difference between cell power and AWGN.
Amplitude offset causing total RF power limits to be exceeded (see
SYSTem:CORRection:GAIN to set or query the amplitude offset value).
Setting Range: 1|ON|0|OFF
Query Range: 1|0
*RST Setting ON
Programming Example
OUTPUT 714;"CALL:CELL:POWER:STATE OFF" !Turns off cell power.

Table of Contents

Other manuals for Agilent Technologies 8960 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 8960 Series and is the answer not in the manual?

Agilent Technologies 8960 Series Specifications

General IconGeneral
BrandAgilent Technologies
Model8960 Series
CategoryTest Equipment
LanguageEnglish

Related product manuals