EasyManua.ls Logo

Agilent Technologies ENA Series

Agilent Technologies ENA Series
769 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
588 Appendix A
Manual Changes
Manual Changes
Figure A-4 Calibration kit definition screen
Step 2. Define each standard as follows
Reference impedance (Z0)
Enter a value of the reference impedance of thru, line and match (1 in Figure A-4).
Reflection
Select a standard type (2 in Figure A-4) and enter a delay value (3 in Figure A-4) in ps.
Thru
Enter a delay value (4 in Figure A-4) in ps. If necessary, also enter its offset loss value (5 in
Figure A-4) in Gohm/sec.
Match
Enter a frequency range (6 in Figure A-4) in GHz. Check the match you use for calibration
(7 in Figure A-4).
Line
You can define 3 lines: Line 1, Line 2, and Line 3.
Enter a delay value (8 in Figure A-4) in ps and enter a frequency range (9 in Figure A-4) in
GHz.
Check the lines you use for calibration (7 in Figure A-4).

Table of Contents

Other manuals for Agilent Technologies ENA Series

Related product manuals