Appendix D 667
Softkey Functions
Calibration Menu
D. Softkey Functions
(Continued)
Calibrate
2-Port Cal
Transmission Displays softkeys for executing transmission calibrations. Symbols x and y refer to the two test ports selected in
the
Select Ports menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
Port x-y Thru Executes a THRU calibration on test ports x and y. :SENS{1-16}:CORR:COLL:THRU x,y
:SENS{1-16}:CORR:COLL:THRU y,x
Return Returns to the softkey display screen one level higher.
Isolation
(Optional)
Displays softkeys for executing isolation calibrations. Symbols x and y refer to the two test ports selected in the
Select Ports menu (x and y are always 1 and 2, respectively, for models with Option 213 or 214).
Port x-y Isol Executes an isolation calibration on test ports x and y. :SENS{1-16}:CORR:COLL:ISOL x,y
:SENS{1-16}:CORR:COLL:ISOL y,x
Return Returns to the softkey display screen one level higher.
Done Terminates a calibration process and calculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:SAVE
Overwrite Terminates partial overwrite and recalculates the calibration
coefficients.
:SENS{1-16}:CORR:COLL:PART:SAVE
Cancel Displays softkeys for canceling a calibration.
OK Cancels a calibration process and clears the measurement data of
standard.
:SENS{1-16}:CORR:COLL:CLE
Cancel Returns to the softkey display screen one level higher.
Return Returns to the softkey display screen one level higher.
3-Port Cal
*1
Displays softkeys for executing full 3-port calibrations.
Select Ports
*2
Displays softkeys for selecting test ports.
1-2-3 Selects test ports 1, 2, and 3.
:SENS{1-16}:CORR:COLL:METH:SOLT3 1,2,3
1-2-4 Selects test ports 1, 2, and 4.
:SENS{1-16}:CORR:COLL:METH:SOLT3 1,2,4
1-3-4 Selects test ports 1, 3, and 4.
:SENS{1-16}:CORR:COLL:METH:SOLT3 1,3,4
2-3-4 Selects test ports 2, 3, and 4.
:SENS{1-16}:CORR:COLL:METH:SOLT3 2,3,4
Cancel Returns to the softkey display screen one level higher.
Reflection Displays softkeys for executing reflection calibrations. Symbols x, y, and z refer to the three test ports selected in
the
Select Ports menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
Port x Open Executes an OPEN reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:OPEN x
Port x Short Executes a SHORT reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:SHOR x
Port x Load Executes a LOAD reflection calibration on test port x.
:SENS{1-16}:CORR:COLL:LOAD x
Port y Open Executes an OPEN reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:OPEN y
Port y Short Executes a SHORT reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:SHOR y
Port y Load Executes a LOAD reflection calibration on test port y.
:SENS{1-16}:CORR:COLL:LOAD y
Port z Open Executes an OPEN reflection calibration on test port z.
:SENS{1-16}:CORR:COLL:OPEN z
Port z Short Executes a SHORT reflection calibration on test port z.
:SENS{1-16}:CORR:COLL:SHOR z
Port z Load Executes a LOAD reflection calibration on test port z.
:SENS{1-16}:CORR:COLL:LOAD z
Return Returns to the softkey display screen one level higher.
Transmission Displays softkeys for executing transmission calibrations. Symbols x, y, and z refer to the three test ports selected
in the
Select Ports menu (x, y, and z are always 1, 2, and 3, respectively, for models with Option 313 or 314).
Port x-y Thru Executes a THRU calibration on test ports x and y. :SENS{1-16}:CORR:COLL:THRU x,y
:SENS{1-16}:CORR:COLL:THRU y,x
Port x-z Thru Executes a THRU calibration on test ports x and z. :SENS{1-16}:CORR:COLL:THRU x,z
:SENS{1-16}:CORR:COLL:THRU z,x
Key Operation Function SCPI Command