272 100-412-234 REV. 06
D.2.10 Recall Preset
The ability to recall a Preset is also available via a web service request. There are only 3
conditions required to recall a Preset: the proper authority is given to the logged in user,
the Preset requested has previously been saved, and the Power Supply is in Ready.
If an attempt is made to recall a Preset when the system is not in Ready, then the
appropriate error code is sent. Simply retry the request if this is the error response
received. If multiple retries fails, then something is preventing the Power Supply from
being Ready and further investigation should be done.
URL
http://<2000Xc IP Address>/Services/RecallPreset
POST Data
{"Sid":12345,"PresetNum":24}
Successful Response
{“ErrorCode”:0,“PresetName”:”Tm 0.25S”}
D.2.11 Verify Preset
Verifying a Preset is necessary before running it. This web service request makes that
possible. All that is needed is the Preset number and the SID. Currently there is no way to
know from a web service request why a Preset verification has failed. One possible reason
is because the Power Supply is not Ready.
Verifying a Preset will also recall the Preset.
URL
http://<2000Xc IP Address>/Services/VerifyPreset
POST Data
{"Sid":12345,"PresetNum":24}
Successful Response
{“ErrorCode”:0}