EasyManua.ls Logo

Bruker BioSpin Solid State NMR - Page 131

Default Icon
328 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
HETCOR with DUMBO, PMLG or w-PMLG, Using Shapes
User Manual Version 002 BRUKER BIOSPIN 131 (327)
sp0 Power level channel 2 for contact pulse.
spnam0 ramp.100 or similar Shape for contact pulse channel f2.
sp1 set to pl13 To match cnst20.
spnam1 lgs-2 To match ppg calculation of in0, in_f1.
p3 2.5 – 3 µsec 90º pulse channel 2 at pl12.
p15 50 -500μs Contact pulse width.
pcpd2 2*p3 SPINAL64 /TPPM decoupling pulse.
cpdprg2 SPINAL64/TPPM15 Decoupling sequence.
F1
1
H indirect
l0 0 Start value 0, incremented during expt.
l3 2- 4 Multiples of FSLG-periods, increment per row.
in_f1 =in0 as calculated Set according to value calculated by ased.
F2
13
C acquisition
d1 2s Recycle delay.
sw 310ppm Sweep width direct dimension.
aq 16-20 msec
masr 10-15 kHz 15 kHz ok at 100 kHz RF-field.
Table 9.2. Processing Parameters for pmlg-HETCOR (as for FSLG on tyrosine-HCl)
Parameter Value Comment
F2 direct dim
13
C
si 2-4 k
wdw QSINE SSB 2, 3 or 5.
ph_mod pk
F1 indirect
1
H
si 256-1048
mc2 STATES-TPPI
wdw QSINE
ssb 3, 5
ph_mod pk
Table 9.1. Acquisition Parameters for pmlg-HETCOR (on tyrosine-HCl)

Table of Contents

Related product manuals