Basic CP-MAS Experiments
User Manual Version 002 BRUKER BIOSPIN 109 (327)
Figure 7.4. CPTOSS243 Experiment on Tyrosine HCl at 6.5 kHz
Figure 7.4. is a CPTOSS243 experiment on tyrosine HCl at 6.5 kHz sample rota-
tion using a 4 mm CPMAS triple resonance probe at 500 MHz with 243 accumu-
lated transients. No spinning sideband residuals can be observed, with a noise
level below 2% peak to peak compared to the highest peak intensity.