Modified W-PMLG
User Manual Version 002 BRUKER BIOSPIN 289 (327)
Digital Mode Acquisition 23.7
Most parameters stay the same as adjusted in analogue mode.
The correction for the scaling factor must be done after acquisition, changing the
status parameter swh by typing s swh and dividing the value by the scaling factor
(about 0.47 for WPMLG and 0.5 for DUMBO).
Table 23.4. Parameters for Digital Mode
Parameter Value Comment
pulprog dumbod2 or wpmlgd2 AV 3 instruments only.
digmod
dspfirm
digital
sharp or medium
aqmod qsim or dqd
swh 50000-10000 Depending on spectral range and o1.