266 (327) BRUKER BIOSPIN User Manual Version 002
Double-CP
Spectral Processing 20.4
Processing parameters
sw ≈200 ppm Sweep width direct dimension, adjust to experimental require-
ments
F1 indirect
15
N (right column)
td 128-.512 Number of real points
sw ≈100-150 ppm Sweep width indirect dimension
Table 20.3. Recommended Processing Parameters for the DCP 2D
Parameter Value Comment
F1 acquisition
13
C (left column)
si 2-4k FT-size
wdw QSINE Squared sine bell
ssb 2-5 Shifted square sine bell, >2: res. enhancement
ph_mod pk Phase correction if needed
F2 indirect
15
N (right column)
si 512-1024 Zero fill
mc2 STATES-TPPI
wdw QSINE Squared sine bell
ssb 2-5
Table 20.2. Recommended Parameters for the DCP 2D Setup