EasyManua.ls Logo

Bruker BioSpin Solid State NMR - 2 D Data Acquisition; Figure 18.6. Signal Intensities of 87 Rb in Rbno3

Default Icon
328 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
238 (327) BRUKER BIOSPIN User Manual Version 002
MQ-MAS: Sensitivity Enhancement
one rotor period and the spinning frequency is increased (and hence the rotor pe-
riod decreased). On “real life” samples the differences between signal intensities
at one rotor period compared to e.g. ¼ rotor periods will be more pronounced than
on a crystalline model compound. Since the spinning frequency is usually deter
-
mined by the spectrum itself, the only degree of freedom is the amplitude of the
sweep. In the example chosen here any of the conditions will provide a good qual
-
ity spectrum, and the condition with the biggest enhancement at the least power is
to be preferred.
Figure 18.6. Signal Intensities of
87
Rb in RbNO
3
Signal intensities of
87
Rb in RbNO
3
as function of duration and RF field amplitude
for double frequency sweeps.
2D Data Acquisition 18.2.2
After the parameters for the DFS are adjusted the 2D data acquisition can be pre-
pared. In tables 2 and 3 the important parameters are listed for the two pulse se-
quences. Parameters are listed separately for F1 and for the pulse program
relevant parameters which should be set in eda and ased, respectively.
The 3-pulse sequence used in mp3qdfs.av creates a phase modulated data set
and therefore, FnMODE must be QF. However, since a whole echo acquisition is
performed a pure absorption mode spectrum can be obtained. Increments for D10
and D11 must be set correctly so that a standard two-dimensional FT can be ap
-
plied.
Using the 4-pulse sequence mp3qdfsz.av FnMode must be States or States-TPPI,
so that the shearing FT can be performed for processing. However, no shearing is
required in case of nuclei with spin I=3/2 where a split-t
1
experiment can be per-
formed, in which case IN10 must be set correctly.

Table of Contents

Related product manuals