EasyManua.ls Logo

Bruker BioSpin Solid State NMR - Possible Approaches for Non-13 C Samples; Table 6.1. Power Conversion Table

Default Icon
328 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Practical CP/MAS Spectroscopy on Spin 1/2 Nuclei
User Manual Version 002 BRUKER BIOSPIN 101 (327)
Possible Approaches for Non-13C Samples 6.3
If an arbitrary X-nucleus of spin ½ is under investigation (quadrupolar spins must
be treated separately), the strategy follows the one described above, if the sample
contains the protons bound to
13
C. In this case, running a
13
C cp/mas spectrum
allows setting and determining all proton parameters (recycle time, contact time)
from the
13
C setup. To run the X-nucleus, cross polarized from protons, one just
needs to set the HH-condition from the known proton RF-field, the spin rate, and
the transmitter power at the NMR-Frequency of the X-nucleus such that the effec
-
tive field at the X-frequency equals the effective field at proton frequency ±spin
rate.
Example: setting the HH-condition for
15
N from known parameters for
13
C-CP/
MAS. The gyro-magnetic ration of
15
N is lower by a factor of 2.5 compared to car-
bon (proton frequency: 400 MHz,
13
C-frequency: 100 MHz,
15
N frequency: 40
MHz). The probe efficiency is about the same for
13
C and
15
N (but not
1
H!), so
one needs about 2.5 times higher RF-voltage for the 15N-contact pulse than for
the
13
C-contact pulse, if the spin rate and the proton RF-field are the same. This is
equivalent to 2.5
2
=6.25 times the power in watts! So if ased shows pl1W =150W
for a well optimized 13C-CP setup,
15
N will require 6.25*150 W= 938 W! This is
far above specs, so the same proton contact power level cannot be used, it needs
to be lowered. The maximum allowed power for a contact pulse on
15
N is 500W.
This means that the proton contact power should be lowered by approximately a
factor of sqrt (938/500) 1.37. Precalculating power levels like this will get the pa
-
rameters close enough to see a cp-signal on a good test sample, so further opti-
mization is possible. See "Test Samples" for suitable test samples.
The most efficient way of precalculating power levels for multi-nuclear spectrosco-
py is the following:
1. Determine the power conversion factor for some nuclei of interest on a suitable
test sample, from the low end to the high end of the probe tuning range. This
means measuring a precise 360° pulse (make sure it is 360°, not 180° or
540°!) and the associated power level. Make a table in your lab notebook as
follows (see
"Appendix"):
Table 6.1. Power Conversion Table
Probe: 4mm Triple
Nucleus Frequency P90 (µs) Rf-field (Khz) Power (W or dB) Remarks
1
H/400.13 2.5 100 100 Low range
19
F/376.3 Not available
15
N/40.5 6.5 38.6 300 Probe in double mode
15
N/40.5 6.5 38.6 500 Probe in triple mode C/N
29
Si/79.5 6 41.7 300 Double mode low range
13
C/100.5 4 62.5 150 Double mode low range
13
C/100.5 5 50 200 Triple mode C/N
119
Sn/149.1 4 62.5 100 Double mode high range
31
P/161.9 3.5 71.4 150 Range switch up, double mode

Table of Contents

Related product manuals