EasyManua.ls Logo

Bruker BioSpin Solid State NMR - T1 P Relaxation Measurements

Default Icon
328 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Relaxation Measurements
User Manual Version 002 BRUKER BIOSPIN 209 (327)
Data Processing
The saturation-recovery data should be processed in the same way as the inver-
sion-recovery data above (see Table 16.3. for parameters). The only differences
are that the fitting function should be satrec rather than expdec, and the slice se-
lected for processing should be the last one (signal is maximum at long recovery
times). The calculated relaxation time constants should be the same as those ob
-
tained by inversion-recovery.
T1p Relaxation Measurements 16.2.5
Rotating-frame relaxation measurements, under a spin-locking rf field, can be
used to probe motions on shorter domestically than T
1
measurements, with in-
verse correlation times of the order of the spin-locking rf field strength.
To measure T
1r
relaxation, after CP a variable length spin-locking pulse is applied
to the X nucleus. The remaining X magnetization decays exponentially to zero, as
a function of spin-lock time. The parameters of cross-polarization can also be de
-
termined from variable-contact-time CP experiments (the function cpt1rho is pro-
vided in the relaxation analysis tool for this purpose), but here only simple T
1r
measurements will be discussed.
It should be noted that the relaxation in a T
1r
experiment might result from pro-
cesses other than true T
1r
relaxation. For example, in glycine, the carbon spins
are dipolar coupled to protons, and there is a possible fast relaxation pathway via
the protons, which is not T
1r
relaxation. This is inhibited by having a high spin-lock
field strength, but at large field strengths care must be taken over the length of the
spin-lock pulse. If apparently non exponential decay is observed, this may result
from such alternative relaxation processes.
Experiment setup
Sample: Glycine
Spinning speed: 10 kHz
Time: 20 minutes
Table 16.4. Parameters for the Saturation Recovery Experiment
Parameter Value Comments
Parmode 2D
Vdlist See text
td(f1) Number of entries in vd list
FnMODE QF This is not a real 2D experiment.
NS 16, for the glycine sample More scans needed than for the CP experiment,
due to reduced signal.

Table of Contents

Related product manuals