Mode
Sets the stabilizer's mode: On, Off, or Hold.
Divider
Sets the count rate dividers at the input to the correction registers, in steps of 2
0
to
2
9
(1 to 512).
Ratio Mode
Sets the stabilizer's correction mode and ratio.
•
In Automatic mode the stabilizer will calculate the ratio factor automatically
when the Reset button is pressed.
•
In Manual mode, the ratio factor must be specified, using the Ratio slidebar
control and display.
Status
Displays the Stabilizer's last reported status.
Correction value
Displays the stabilizer's last reported correction value.
Reset button
The Reset button (located in the Status field, not the button at the bottom of the
webpage) will apply any new Stabilizer settings, based on current settings.
Pressing the Reset button will cause any new Stabilizer settings to be applied, and it
may require a few seconds before the results will be updated to the display.