Readings and Faults Table
The results of the last diagnostic test will be displayed unless the Clear faults Exe
-
cute button has been pressed.
If any test returns a failure status for the selected test routines, the results will be dis
-
played as a 'Fault' in the Faults table. Power Supply results are indicated in the Voltage
Readings table.
Voltage Readings
Displays the current voltage levels and status of the Low Voltage Power Supply
(source, voltage reading, and status). If any are out of tolerance, a fault is annunciated
and the row is highlighted in red.
Inputs
The Inputs table reports the status of each of the Inputs, and if a User has locked an in
-
put, that User ID is listed.
Faults
A Faults log keeps track of all faults recorded by the system, including the faults gen-
erated by the Diagnostics command.
Tabulates all faults (or 'No faults detected' status) that were detected the last time the
Diagnostics were run, unless the Faults have been cleared. The fault descriptions, or
no faults, will be listed under the Description heading in the Faults area.
User’s Manual 139
Diagnostics | Faults Maintenance