Stabilizer
The Stabilizer maintains the stability of high resolution spectroscopy in applications
involving long count times or high count rates.
The Stabilizer settings may be adjusted and will be applied using the Reset button in
the Status window.
Spacing
Sets the number of channels between the upper and lower sampling windows.
Window
Sets the width (in channels) of the upper and lower sampling windows on either side
of the gain reference peak.
Ratio
The Ratio field shows the current setting for the Ratio factor, in a range from 0.010 to
100.0. The slidebar control sets the stabilizer ratio factor for the Manual mode.
Centroid
Sets the value (in channels) of the reference peak for gain stabilization.
Range
Sets the correction range to correct for drift, for either Ge or NaI detector types. .
User’s Manual 89
Stabilizer | Range MCA Settings
Figure 68 - Stabilizer Setup Page