HVPS Range 2
The following applies to HVPS 2:
OUTPUT – ±1500–5000V at 1 µA max.
RIPPLE – 10 mV P-P.
TEMPERATURE COEFFICIENT – ±50 ppm/ºC.
STABILITY – 0.01%/h, 0.02%/8 h.
ACCURACY – ±2%.
LOAD REGULATION – 1%.
SETTING RESOLUTION – 12-bit (1/4096).
HVPS Range 3
The following applies to HVPS 3:
OUTPUT – ±200 V at 100 nA max.
RIPPLE – 10 mV P-P.
TEMPERATURE COEFFICIENT – ±50 ppm/ºC.
STABILITY – 0.01%/h, 0.02%/8 h.
ACCURACY – ±5%.
LOAD REGULATION – 1%.
SETTING RESOLUTION – 12-bit (1/4096).
Physical
MAIN ENCLOSURE – Metal
DIMENSIONS – 9.0 x 21.3 x 27.4 cm (3.5 x 8.4 x 10.8 in.) (H x W x D).
For rack-mounting purposes, this is one-half-width of a standard 19-inch electronics
cabinet wide by 2U high; two units may be configured for side-by-side rack mount
orientation using the available rack-mounting kit.
WEIGHT – 2.6 kg (5.7 lb).
User’s Manual 155