HV
Green LED that indicates the state of the selected high voltage power supply:
•
Off – High voltage supply is off.
•
Green Blinking – High voltage supply is ON and ramping to its programmed
voltage.
•
Green Steady On – High voltage supply is ON and has reached its programmed
voltage.
ICR
Green LED that illuminates each time the instrument processes a nuclear event. The
intensity of the LED will vary in proportion to the incoming count rate.
Performance
Signal Processing
GAIN – x2.4 to x430 in increments of ½16.
FILTER – Trapezoidal.
SHAPING TIMES – Rise time: 0.32 to 51.2 µs; Flat top: 0 to 3.2 µs.
AUTO POLE/ZERO – 40 µs to infinity.
BASE LINE RESTORER – Automatic with provision for adjustable BLR.
LLD SETTING – 0 to 100%.
ULD SETTING – 0 to 100%.
INHIBIT MODE – Programmable 0 to 5 ms.
LT TRIM – 0 to 1000.
PUR GUARD – Programmable guard time 0 to 2.5 times shaping.
Appendix A - Specifications
150 Lynx™ Digital Signal Analyzer