LED Status
POWER
A multicolor LED showing instrument status and presence of system power. The dif
-
ferent LED states may be observed during or after the power-up self test.
Blinking Green – Instrument is powered on and in the process of initialization.
Steady Green – Instrument is powered on and ready for operation.
Steady Red – Signals a self-diagnostic error. The reason for the error can be
confirmed through the appropriate control software (Genie 2000 or the web
interface). A Diagnostic procedure is available if the Lynx interface is
operational.
LED is Off – Instrument is not receiving power.
ACQUIRE
A green LED. Steady on when a setup on the instrument is acquiring data.
COMM
A green LED for Communications. Illuminates each time the instrument has received
a command over any supported communications interface.
HV
A multicolor LED indicating the state of the selected high voltage power supply, or in
some cases, relating to any of the high voltage power supply outputs.
Off – High voltage supply is off. ALL HVPS Outputs are off.
Blinking Green – ANY High voltage supply is ON and ramping up to its
programmed voltage.
Steady On Green – ANY High voltage supply is ON and has reached its
programmed voltage.
Steady On Red - An error, such as a HVPS fault has occurred (e.g., Inhibit has
tripped, etc.). This 'error' condition may be cleared by the User with knowledge
of what caused the fault to occur (i.e., a known limiter was tripped).
ICR
A green LED for Incoming Count Rate. Illuminates each time the instrument processes
an incoming signal. The intensity of the LED is proportional to the incoming count
rate.