Factory Reset Control
The factory reset button is located behind a small hole in the Lynx's bottom panel.
CAUTION Operating this control will cause a loss of stored data, and will
interrupt an acquisition in process.
To reset the instrument to a factory default condition, with power applied to the Lynx,
press and hold the button with a small screwdriver as noted below:
•
Press and hold for 1 second to reset the instrument to the factory default
condition and settings.
•
Press and hold for 7 seconds to perform a soft reboot of the instrument.
Chapter 2 - Controls and Connectors
22 Lynx™ Digital Signal Analyzer