Timeout
Defines the External Sync Timeout duration, in milliseconds. A default of 200 mil
-
liseconds is set. The timeout is related to the amount of time to wait for the pulse be
-
fore timing out.
Polarity
The signal polarity required for the external sync signal.
Status
This feature provides the ability to Enable or Disable synchronization, and to select
an Independent external control option via the SYNC input connector if desired.
Independent status allows independent synchronization control for SCA, Auxiliary
Counters, MSS, and List data collection. If Independent is selected, the selection for
Internal clock or External control using the SYNC input is found on the Acquisition
Setup page. See MCA Settings | Acquisition Setup for more information.
Chapter 4 - Web-based Operation External Synchronization