10.2 Synchronized Measurement
187
Chapter 10 Setting Method for Logger
10
This section describes setting synchronized measurement to MASTER or
SLAVE when you want to measure by synchronizing the sampling of multiple
loggers.
1. Turn the power on while pressing
and then release the key
when the model name and version
appear in the display area.
The instrument enters setting mode.
2. Press multiple times until the following synchronized measurement
setting screen appears.
3. Press to select the synchronization setting.
Each press of the key toggles the selection item.
4. Press to move the flashing cursor to [SET], and press to con-
firm the setting.
5. Press multiple times until the following setting mode end screen
appears.
6. Press to end setting mode.
10.2 Synchronized Measurement
Selectable
Items
Explanations
MASTER Set this instrument as the master and reference for sampling.
SLAVE
Set this instrument as the slave and match the reference for
sampling to the master.
OFF Do not set synchronized measurement.
Multiple masters cannot be assigned to one measurement system.