Appendix 3 Error Display
A6
Warning Messages
When an error occurs, a message appears in the display area and flashes red. The warning message
disappears when is pressed.
No. Messages Meaning Solution
201 Can'tMoveDrtory The data could not be moved to
the directory used last time or the
specified directory.
When saving data to a directory other than the
standard \LOGGER directory, use a CF card on
which that directory has already been created.
202 File locked An attempt was made to write to a
read-only file, or an attempt is
being made to read/write from or to
a file currently being used for
another task.
Cancel the read-only attribute of the file, or wait
until the other task is finished.
203 Read only file The file cannot be modified or
deleted because it is a read-only
file.
Cancel the read-only attribute.
204 Bad path or name There is an error with the path
name or file name.
Use the correct path name or file name.
205 Same File exists A file of the same name exists. Change the file name.
207 Path too long The path name is too long. Make sure the path name is no more than 127
characters long.
208 Set CardDir Name There was no name when
attempting to create a directory.
Specify the directory name.
209 Same Dir. Exists A directory of the same name
exists.
Change the directory name.
212 Bad file data An attempt was made to read a file
of a different format.
If the file was saved by an instrument of a version
newer than that of the instrument, update the
version of the instrument. This message also
appears if an attempt is made to read a damaged
file or a file of another format.
213 program file err The file for updating the version
could not be found.
Make sure there is only one 8423 version update
file in the root directory of the CF card.
214 setup file err The file for setting the instrument
could not be found.
Make sure there is only one 8423 settings file in
the root directory of the CF card.
226 Card Almost full There is insufficient space on the
CF card.
This message appears when the available space
is 10% or less, or 16 MB or less. Replace the CF
card with a new formatted one if necessary.
227 Old file deleted An old file was deleted with the
delete and save function.
An old file was deleted because the delete and
save setting is set. Replace the CF card with a
new formatted one if necessary.
303 No wave data There is no waveform data in the
memory of the instrument.
Start measurement so that the instrument can
obtain waveform data.
520 Bad MAC address The MAC address of the
instrument has been altered.
Contact your dealer or Hioki representative.
521 Bad IP address The IP address is invalid. Check the IP address setting.
527 Comm. Error A communication error occurred. Check the communication settings of the
instrument, communication cable,
communication circuit, and communication
destination.
529 DHCP failure An IP address could not be
obtained with DHCP.
Check the communication settings of the
instrument, communication cable,
communication circuit, and DHCP server.

Table of Contents