EasyManua.ls Logo

Hioki 8423 - Page 59

Hioki 8423
266 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
3.8 Connection Method for External Control Input Terminals
51
7
3
Chapter 3 Setting Up the Instrument
Required tools: Flat blade screwdriver............... 1
1. Press the button of the terminal with a flat blade screwdriver (or other
tool).
2. While keeping the button pressed in, insert the wire into the hole of the ter-
minal.
3. Release the button to lock the wire into place.
<Circuit Diagram of External Control Input Terminals>
* 10 μF when the chattering prevention filter is on.
*
When using multiple loggers to measure with the external sampling function,
care should be taken with the following points.
Always connect the 9683 Connection Cables when synchronized measure-
ment will be performed. Furthermore, only use the SMPL/TRIG terminal for the
master.
Disconnect the 9683 Connection Cables when synchronized measurement will
not be performed. Furthermore, input the same signal into the SMPL/TRIG ter-
minal of all loggers.

Table of Contents

Related product manuals