8: TSP command reference Model 2461 Interactive SourceMeter® Instrument
Reference Manual
8-336 2461-901-01 A/November 2015
Details
The tspnet.write() function sends inputString to the remote instrument. It does not wait for
command completion on the remote instrument.
The Model 2461 sends inputString to the remote instrument exactly as indicated. The
inputString must contain any necessary new lines, termination, or other syntax elements needed
to complete properly.
Because tspnet.write() does not process output from the remote instrument, do not send
commands that generate too much output without processing the output. This command can stop
executing if there is too much unprocessed output from previous commands.
Example
tspnet.write(myID, "runscript()\r\n")
Commands the remote instrument to execute
a command or script named runscript() on
a remote device identified in the system as
myID
.
Also see
tspnet.connect() (on page 8-325)
tspnet.read() (on page 8-328)
upgrade.previous()
This function returns to a previous version of the Model 2461 firmware.
Type TSP-Link accessible
Affected by Where saved Default value
Function No
Usage
upgrade.previous()
Details
This function allows you to revert to an earlier version of the firmware.
When you send this function, the instrument searches the USB flash drive that is inserted in the
front-panel USB port for an upgrade file. If the file is found, the instrument performs the upgrade. An
error is returned if an upgrade file is not found.
Use this command with caution. Make sure your instrument can support the earlier version and that
there are no compatibility issues. Check with Keithley Instruments before using this command if you
have questions.
Also see
Upgrading the firmware (on page A-3)
upgrade.unit() (on page 8-337)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV