Model 2461
Interactive SourceMeter® Instrument Reference Manual Section 8:
TSP command reference
2461-901-01 A/November 2015 8-137
smu.measure.configlist.delete()
This function deletes a measure configuration list.
Type TSP-Link accessible
Affected by Where saved Default value
Function
Yes
Usage
smu.measure.configlist.delete(listName)
smu.measure.configlist.delete(listName, index)
listName
A string that represents the name of a measure configuration list
index
A number that defines a specific configuration index in the configuration list
Details
Deletes a configuration list. If the index is not specified, the entire configuration list is deleted. If the
index is specified, only the specified configuration index in the list is deleted.
When an index is deleted from a configuration list, the index numbers of the following indexes are
shifted up by one. For example, if you have a configuration list with 10 indexes and you delete index
3, the index that was numbered 4 becomes index 3, and the all the following indexes are renumbered
in sequence to index 9. Because of this, if you want to delete several nonconsecutive indexes in a
configuration list, it is best to delete the higher numbered index first, then the next lower index, and so
on. This also means that if you want to delete all the indexes in a configuration list, you must delete
index 1 repeatedly until all indexes have been removed.
Example
smu.measure.configlist.delete("myMeasList")
Delete a measure configuration list named
myMeasList.
smu.measure.configlist.delete("myMeasList", 2)
Delete configuration index 2 from the measure
configuration list named myMeasList.
Also see
Configuration lists (on page 3-30)
smu.measure.configlist.create() (on page 8-136)

Table of Contents

Other manuals for Keithley 2461

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Keithley 2461 and is the answer not in the manual?

Keithley 2461 Specifications

General IconGeneral
Maximum Current7 A
Power100 W
Resolution6.5 digits
Source Measure Unit (SMU)Yes
DC Voltage100 V
Maximum Voltage100 V
Display5-inch capacitive touchscreen
ConnectivityUSB, GPIB
SoftwareTSP
DC Current7 A
Current Source Range0 to 7 A
Voltage Measurement Range100 mV to 100 V
Current Measurement Range10 nA to 7 A
Voltage Measurement Resolution100 nV

Related product manuals